Scheme for Determination of Methanol Content in Liquor by Gas Chromatography
Beijing Huasheng Spectrum Instrument Co., Ltd.
First, the purpose of the experiment
1.
2.
FID
Performance and method of operation.
3.
4.
5 g
Serious poisoning will occur, exceeding12. 5 g
It may lead to death. In the fermentation process of liquor, it is difficult to completely separate methanol and ethanol. Therefore, the state has strict regulations on the methanol content of liquor. According to national standards (GB10343-89
), the methanol content of edible alcohol should be lower than0.1g
•L-1 (
Gifted class)
or0.6 g
•L-1
(normal level).
1.
Gas Chromatograph
, hydrogen flame detector (FID
), chromatography workstation, capillary column,10μL
Micro syringe1
support,100mL
Volumetric flask1
only,25mL
Volumetric flask5
only,5mL
Pipette1
only.
2.
:
Methanol (chromatographic alcohol),60%
Aqueous ethanol solution (without methanol). take1
μL
Inject no methanol peak.
1.
0.32mm
Long30m
Liquid film thickness0.5μm
The quartz capillary column is installed and debugged normally.
2.
FID
Inlet temperature180
°C; detector temperature200
°C; column temperature50
°C.
3.
100mL
60%
Ethanol solution,
Then weigh accurately on the electronic balance0.5000g
Chromatography of methanol in this volumetric flask,
Last use60%
Constant volume of ethanol solution,
This solution is5g/L
Methanol stock solution.1.0
,2.0
,3.0
,4.0
,5.0mL,
Placed separately5
One25mL
Volumetric flask,
use60%
Diluted ethanol to the scale,
Mix and get separately0.2
,0.4
,0.6
,0.8
,1.0 mg/mL
Methanol standard solution series,
.
4.
1
) Turn on the gas generator and confirm that the on/off valves on the purifier are in the fully open position (ON
), wait until the pressure of nitrogen, hydrogen, and air reaches the corresponding value (about30-40min
), turn on the instrument power switch, enter the workstation, set the inlet temperature180
°C, detector temperature200
°C, column temperature50
°C. Click on the corresponding detector signal channel (FID
:1
Channel), and other instruments are stable.2
Ignition: Check the detector temperature. After the detector temperature reaches the set value, confirm that the air and hydrogen valves of the corresponding detector on the gas pavement on the right side of the instrument are in the fully open position. Press the ignition key to confirm that the detector fire used is lit (see if there is water vapor generation and baseline changes). After the baseline level, the samples were injected in turn.3
After the experiment is finished, the hydrogen generator should be turned off first, and then the oven temperature is set to30
Degree, the injector and detector are turned off, the temperature is lowered, and the oven temperature is lower than50
Below °C to exit the software, turn off the instrument power switch, turn off the gas generator.
4.
1μL
Each chromatographic grade of methanol was injected into the chromatograph to obtain a chromatogram, and the retention time was used as a control to determine the methanol peak.,
Accurately take the syringe from low to high1
μL
Standard use of liquid for analysis,
Each concentration point is analyzed by a chromatograph3
Times, measured their peak area,
The result is the average peak area of ​​methanol as the ordinate.Y
, methanol concentration is the abscissax
, draw a standard working curve, get linear equations and correlation coefficients.
5.
,
draw1μL
Liquor samples were analyzed and the results were directly calculated by the chromatography workstation. Sample repeat measurement3
Times,
take the average,
The content of methanol in the liquor sample can be obtained.Power Film Capacitors - Plastic Case
Polypropylene High Voltage Capacitor,axial film capacitor,axial capacitor,SHV capacitor
XIAN STATE IMPORT & EXPORT CORP. , https://www.shvcomponents.com